Baxter Infus OR
This product is also in these categories:
Product Information:
- Products Model: 70231IRS
- Product Brand: DRE Medical
- Condition: Factory Refurbished
- Warranty: Satisfaction Guaranteed
The Baxter Bard Infus O.R. Syringe Pump automates drug delivery calculations using magnetically coded SMART LABELs that configure infusion parameters without manual programming. When you attach a labeled syringe, the pump reads the magnetic code and sets the appropriate flow rates, concentration limits, and dosing protocols for that specific medication. This eliminates the calculation errors that occur with manual parameter entry, particularly critical when working with vasoactive drugs or anesthetics that have narrow therapeutic windows.
Real-time dose adjustments happen through a single-hand dial control while the infusion continues running. You can titrate medications up or down based on patient response without stopping the pump or going through reprogramming sequences. The compact design fits easily on cramped OR tables and runs on standard alkaline batteries, so you won’t lose settings during patient transport or power interruptions.
Key Features
- SMART LABEL magnetic coding system eliminates manual drug calculations
- Single-hand dial control for real-time dose titration during active infusion
- Battery operation with four C-cell alkalines for transport reliability
- Accepts standard 20cc Monoject and B-D syringes
- Compact 9.2″ x 4.5″ x 2.0″ footprint, weighs 2 lbs with batteries
- Stainless steel construction for OR environment durability
Specifications
| Contents | |
|---|---|
| Family/Type | set |
| Sizes | |
| Options | 0; 1; 2; 3; 4; newborn |
| Materials | |
| Components | Stainless Steel |
| Syringes | |
| Compatible Syringes | Monoject® or B-D®; 20 cc |
| Physical | |
| Height | 9.2 inches |
| Width | 4.5 inches |
| Depth | 2.0 inches |
| Weight | 2 lbs (with batteries) |
| Power | |
| Batteries | Four “C” alkaline cells |
| Operating Voltage | 5 – 7 volts |
| Low Battery Voltage | 5.0 volts (approx.) |
| Battery Life | Typical Use: 150 hours; at low battery and typical use: 4 hours |
| Performance | |
| Flow Rates | 0 – 600 mL/hr |
| Flow Profile | Bolus: Continuous; Infusion: Pulsed Intermittently |
| Accuracy | Infusion: linear rate ±3%; Bolus: linear displacement ±3% |
| Occlusion Force | 8 ± 1 lb. |
| Maximum Occlusion Pressure | 11 psi with 60 cc syringe; 20 psi with 20 cc syringe |
| Occlusion Detection Time | 120 seconds at 36 mL/hr (inverse related to flow rate) |
| Volume Stored on Occlusion | 1.1 mL (approx.) |
| Back Pressure Effect on Accuracy | None to Occlusion Pressure |
| Intended Use | |
| Primary Use | Administration of narcotics, muscle relaxants, and vasoactive drugs during anesthetic and outpatient surgical settings |
| Illumination | |
| Illumination Types | conventional; fiber optic |